Key result
Non-invasive blood pressure measurement via brachial cuff oscillometry showed large limits of agreement compared to invasive measurement, with a mean difference for MAP of -1.0 mmHg.
Why the study?
The study aimed to compare non-invasive blood pressure measurements with invasive arterial blood pressure measurements in critically ill patients.
Does non-invasive blood pressure measurement using brachial cuff oscillometry provide accurate readings compared to invasive arterial blood pressure measurement in critically ill patients?
Observational (n=736)
Does non-invasive blood pressure measurement using brachial cuff oscillometry provide accurate readings compared to invasive arterial blood pressure measurement in critically ill patients?
Non-invasive brachial cuff oscillometry shows large limits of agreement compared to invasive arterial measurements in critically ill patients, potentially leading to different treatment decisions in about 20% of cases.
Answers are grounded in this paper and cite their sources.
May prompt selective invasive monitoring in critically ill patients; leaves open validation of alternative non-invasive methods.
PURPOSE: The aim was to compare non-invasive blood pressure measurements with invasive blood pressure measurements in critically ill patients. METHODS: Non-invasive blood pressure was measured via automated brachial cuff oscillometry, and simultaneously the radial arterial catheter-derived measurement was recorded as part of a prospective observational study. Measurements of systolic arterial pressure (SAP), diastolic arterial pressure (DAP), and mean arterial pressure (MAP) were compared using Bland-Altman and error grid analyses. RESULTS: Paired measurements of blood pressure were available for 736 patients. Observed mean difference (±SD, 95% limits of agreement) between oscillometrically and invasively measured blood pressure was 0.8 mmHg (±15.7 mmHg, -30.2 to 31.7 mmHg) for SAP, -2.9 mmHg (±11.0 mmHg, -24.5 to 18.6 mmHg) for DAP, and -1.0 mmHg (±10.2 mmHg, -21.0 to 18.9 mmHg) for MAP. Error grid analysis showed that the proportions of measurements in risk zones A to E were 78.3%, 20.7%, 1.0%, 0%, and 0.1% for MAP. CONCLUSION: Non-invasive blood pressure measurements using brachial cuff oscillometry showed large limits of agreement compared to invasive measurements in critically ill patients. Error grid analysis showed that measurement differences between oscillometry and the arterial catheter would potentially have triggered at least low-risk treatment decisions in one in five patients.
No takes yet. Share an insight, caveat, or question.
Kaufmann et al. (2020) conducted an observational in Critically ill (n=736). Non-invasive automated brachial cuff oscillometry vs. Invasive radial arterial catheter-derived measurement was evaluated on Mean difference in mean arterial pressure (MAP) (MD -1.0 mmHg, 95% CI -21.0 to 18.9). Non-invasive blood pressure measurement via brachial cuff oscillometry showed large limits of agreement compared to invasive measurement, with a mean difference for MAP of -1.0 mmHg.
Synapse has enriched 10 related papers. Consider them for comparative context: