Key result
A two-input model integrating physical activity tracking reduced continuous glucose monitoring error during physical activity from 17.46% to 13.8%.
Why the study?
Current continuous glucose monitors exhibit increased estimation error during periods of aerobic physical activity.
Does the integration of physical activity tracking signals reduce Continuous Glucose Monitor (CGM) estimation error during exercise?
Does the integration of physical activity tracking signals reduce Continuous Glucose Monitor (CGM) estimation error during exercise?
Integrating physical activity data from wearable devices into CGM algorithms can significantly reduce glucose estimation errors during exercise.
Answers are grounded in this paper and cite their sources.
Supports refining CGM algorithms with activity data during exercise; hypothesis-generating and should not yet change practice.
Current Continuous Glucose Monitors (CGM) exhibit increased estimation error during periods of aerobic physical activity. The use of readily-available exercise monitoring devices opens new possibilities for accuracy enhancement during these periods. The viability of an array of physical activity signals provided by three different wearable devices was considered. Linear regression models were used in this work to evaluate the correction capabilities of each of the wearable signals and propose a model for CGM correction during exercise. A simple two-input model can reduce CGM error during physical activity (17.46% vs. 13.8%, p < 0.005) to the magnitude of the baseline error level (13.61%). The CGM error is not worsened in periods without physical activity. The signals identified as optimal inputs for the model are “Mets” (Metabolic Equivalent of Tasks) from the Fitbit Charge HR device, which is a normalized measurement of energy expenditure, and the skin temperature reading provided by the Microsoft Band 2 device. A simpler one-input model using only “Mets” is also viable for a more immediate implementation of this correction into market devices.
No takes yet. Share an insight, caveat, or question.
Sanz et al. (2019) studied this question. Two-input model integrating physical activity tracking vs. Standard continuous glucose monitoring was evaluated on Continuous glucose monitoring error during physical activity (p=< 0.005). A two-input model integrating physical activity tracking reduced continuous glucose monitoring error during physical activity from 17.46% to 13.8%.