A reflective fiber-optic surface plasmon resonance (RFSPR) sensor was proposed for detection of cadmium ion (Cd2+) in water based on polyvinyl alcohol/titanium dioxide (PVA/TiO2) composite sensing film. The silver (Ag) was deposited on the side and end surfaces of plastic cladding multimode fiber (PCMF), and then uniform PVA/TiO2was coated on the outside surface of Ag covered PCMF. The Ag films on the end face and the side face of the fiber are respectively playing the roles in enhancing light reflection and exciting SPR effectively. The Ag-PVA/TiO2composite film is served as the sensing film of Cd2+. The results show that the sensor can detect 0-1~μ MCd2+in aqueous solution. The maximum sensitivity of Cd2+is 315.2 nm/μM, and the minimum detection concentration is0.006~μM (about 0.76 ppb), which is much less than that (3 ppb) of the international standard. This assay for Cd2+detection possesses high sensitivity, excellent selectivity and good stability. The fiber- optic method might become a potential Cd2+detection tool in aqueous solution.
No takes yet. Share an insight, caveat, or question.
Li et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: