Key result
The Colin CBM-7000 tonometric monitor showed limited agreement with invasive measurements, with approximately 34% of mean arterial pressures differing by over 10 mmHg.
Why the study?
Does the Colin CBM-7000 tonometric noninvasive arterial blood pressure monitor provide accurate measurements compared to invasive radial artery monitoring in neuro-intensive care patients?
Observational (n=15)
No
Does the Colin CBM-7000 tonometric noninvasive arterial blood pressure monitor provide accurate measurements compared to invasive radial artery monitoring in neuro-intensive care patients?
The Colin CBM-7000 tonometric noninvasive blood pressure monitor showed limited agreement with invasive measurements and is not recommended for continuous monitoring in the intensive care setting.
Answers are grounded in this paper and cite their sources.
May not support Colin CBM-7000 for continuous neuro-ICU monitoring; leaves open need for rigorous validation of tonometric devices.
Intra-arterial measurement is considered the gold standard for continuous, beat-to-beat arterial blood pressure monitoring. However, arterial cannulation can be difficult and may cause complications such as thrombosis and ischaemia. Recently, a tonometric system, the Colin CBM-7000 has been developed for noninvasive beat-to-beat measurement of arterial blood pressure from the radial artery. We assessed the level of agreement between the CBM-7000 and invasive radial artery measurements in 15 patients on a neuro-intensive care unit. Agreement of systolic, diastolic and mean arterial pressure values was limited, with approximately 34% of mean arterial pressures differing by over 10 mmHg. In many cases, this was due to a downward drift of the noninvasive measurements over time. Furthermore, there was a tendency to underestimate low pressures and overestimate high pressures. In our opinion, the Colin CBM-7000 cannot be recommended for continuous blood pressure monitoring in the intensive care setting.
No takes yet. Share an insight, caveat, or question.
Steiner et al. (2003) conducted an observational in Neuro-intensive care (n=15). Colin CBM-7000 tonometric system vs. Invasive radial artery measurements was evaluated on Agreement of systolic, diastolic and mean arterial pressure values. The Colin CBM-7000 tonometric monitor showed limited agreement with invasive measurements, with approximately 34% of mean arterial pressures differing by over 10 mmHg.