Key result
Before in-vivo calibration, the CDI 500 without gas calibration was completely unreliable compared to the gas-calibrated CDI, though after in-vivo calibration there was no statistical difference.
Why the study?
Does using the CDI 500 in-line blood gas monitor without prior gas calibration reduce the accuracy and precision of blood gas measurements in adult patients undergoing cardiopulmonary bypass?
RCT (n=58)
Randomized
Does using the CDI 500 in-line blood gas monitor without prior gas calibration reduce the accuracy and precision of blood gas measurements in adult patients undergoing cardiopulmonary bypass?
Omitting gas calibration of the CDI 500 in-line blood gas monitor prior to cardiopulmonary bypass results in highly inaccurate initial blood gas readings until an in-vivo calibration can be performed.
Clinicians must gas-calibrate the CDI 500 before bypass to avoid unreliable initial readings; RCT confirms calibration necessity until in-vivo adjustment.
During cardiopulmonary bypass blood gases can be analyzed with laboratory equipment or with an in-line monitor giving instant results. The manufacturer of the CDI 500 in-line blood gas monitor recommends gas calibration before use. In acute cases there may not be time to perform a gas calibration. We hypothesized that after calibration against laboratory results, the CDI values of pH, pO2, and pCO2 will keep the same level of accuracy, whether the CDI has been gas calibrated or not. We performed a prospective randomized observational study using a study group without gas calibration (29 patients) and a control group with gas calibration (29 patients). Blood sampling was done at the beginning of bypass, and 30 minutes later. After each blood sample the CDI was in-vivo calibrated to the values simultaneously obtained from the ABL. Before in-vivo calibration values from the CDI without gas calibration were significantly different from the ABL-values in accuracy as well as precision, whereas the results from the gas calibrated CDI were largely consistent with the ABL. Before in-vivo calibration, the CDI without gas calibration was completely unreliable. After in-vivo calibration there was no statistical difference between the values of the CDI with and without calibration. We recommend gas calibration of the CDI before use in the period before in-vivo calibration.
No takes yet. Share an insight, caveat, or question.
Bellaiche et al. (2011) conducted an RCT in Cardiopulmonary bypass (n=58). CDI 500 in-line blood gas monitor without gas calibration vs. CDI 500 in-line blood gas monitor with gas calibration was evaluated on Accuracy and precision of pH, pO2, and pCO2 values compared to laboratory results. Before in-vivo calibration, the CDI 500 without gas calibration was completely unreliable compared to the gas-calibrated CDI, though after in-vivo calibration there was no statistical difference.