Optimisation of the activation parameters for ion trap mass spectrometric analysis of the chlormequat cation using simplex optimisation enabled the product ion (m/z 58) response to be improved 1000-fold. A comparison of the sensitivity of the optimised ion trap mass spectrometer with that of a triple quadrupole mass spectrometer for liquid chromatography/tandem mass spectrometry (LC/MS/MS) showed that similar limits of detection (LODs) could be achieved. For the MS/MS transition of the (35)Cl precursor to the most abundant product, LODs were 0.8 ng cation mL(-1) (0.004 mg cation kg(-1) pear equivalent) and 1.0 ng cation mL(-1) (0.005 mg cation kg(-1) pear equivalent) on the triple quadrupole and ion trap instrument, respectively.