A new method of single-photon timing and its application to measuring fast fluorescence decays is discussed. The detector-signal processor, having a transient response of 0.77 nsec FWHM, is designed around an inexpensive, two-channel sampling oscilloscope. Using a mode-locked, pulse-picked argon ion laser as an excitation source, the system performance in measuring fast fluorescence decays was evaluated. An uncertainty for lifetimes measured in the range of 4 nsec to 50 psec was found to be ±6% or ±10 psec, whichever is greater.