Answers are grounded in this paper and cite their sources.
The programmable array microscope (PAM) uses a spatial light modulator (SLM) to generate an arbitrary pattern of conjugate illumination and detection elements. The SLM dissects the fluorescent light imaged by the objective into a focal conjugate image, Ic, formed by the ‘in‐focus’ light, and a nonconjugate image, Inc, formed by the ‘out‐of‐focus’ light. We discuss two different schemes for confocal imaging using the PAM. In the first, a grid of points is shifted to scan the complete image. The second, faster approach, uses a short tiled pseudorandom sequence of two‐dimensional patterns. In the first case, Ic is analogous to a confocal image and Inc to a conventional image minus Ic. In the second case Ic and Inc are the sum and the difference, respectively, of a conventional and a confocal image. The pseudorandom sequence approach requires post‐processing to retrieve the confocal part, but generates significantly higher signal levels for an equivalent integration time.
No takes yet. Share an insight, caveat, or question.
Verveer et al. (1998) studied this question.