Key result
Consistent radiographic positioning to within 4 cm on successive imaging studies limits artifactual stent-graft migration errors to 1.5 mm, and true migration can be calculated to within 2 mm.
Artifactual migration of aortic stent-grafts due to radiographic centering variation is usually not clinically significant if centered consistently, and can be mathematically corrected.
May support consistent centering in EVAR surveillance to curb artifactual migration error; leaves open prospective validation of standardized protocols.
Purpose: To demonstrate the influence of radiographic positioning on the assessment of stent-graft migration using plain radiographs following endovascular abdominal aortic aneurysm repair. Methods: Equations were derived to correct for artifactual stent-graft migration introduced by geometric distortion due to variations in positioning between radiographs acquired at different times. A phantom system was used to validate the equations. Results: Errors in stent position increase with (1) the distance of the aortic stent-graft from the midline and (2) differences in radiographic centering points in the craniocaudal direction; other variables have little effect. For typical stent positions, errors are small if the centering changes by <8 cm. Consistent radiographic positioning to within 4 cm on successive imaging studies limits errors to 1.5 mm. Even if artifactual migration is large, the true migration can be reliably calculated to within 2 mm. Conclusions: Artifactual migration due to variation in radiographic centering is not usually clinically significant if care is taken to center radiographs consistently. Radiographs in which artifactual migration may be important are readily identified, and mathematical correction is straightforward.
No takes yet. Share an insight, caveat, or question.
Hodgson et al. (2003) studied Endovascular abdominal aortic aneurysm repair. Radiographic positioning variation was evaluated on Errors in stent position. Consistent radiographic positioning to within 4 cm on successive imaging studies limits artifactual stent-graft migration errors to 1.5 mm, and true migration can be calculated to within 2 mm.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: