Answers are grounded in this paper and cite their sources.
The preparation of solid reagent, (5,7-dichloroquinoline-8-ol) modified naphthalene for preconcentration of uranium is described. The uranium-5,7-dichloroquinoline-8-ol complex is quantitatively retained on naphthalene in the pH range 4.5–7.0. For the preconcentration of uranium an aliquot of the above reagent is added to the uranium sample solution, adjusted to pH 5.5 ± 1.0 and the residue is filtered off and dissolved in acetone for analysis by the arsenazoIII method. Calibration graphs are linear over the uranium concentration range 2–100 µg per 5 ml of final solution. Ten replicate determinations of 40 µg of uranium present in one liter of sample gave a mean absorbance of 0.185 with a relative standard deviation of 1.5%. The detection limit corresponding to 3 times the standard deviation of the blank was found to be 2 ng/ml. The validation of the developed preconcentration procedure was carried out by successfully analysing standard marine sediment reference material.
No takes yet. Share an insight, caveat, or question.
T. Prasada Rao (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: