An impeller design with 400° of blade twist generated measurable pressure rises of 7-28 mmHg for flow rates of 1-4 L/min at 6000-8000 RPM in hydraulic testing.
A novel impeller design with a 400° blade twist angle for an intravascular cavopulmonary assist device generated sufficient pressure rises in computational and hydraulic testing, supporting its continued development for Fontan patients.
Answers are grounded in this paper and cite their sources.
Supports preclinical impeller development for Fontan assist; leaves open clinical translation pending in vivo validation.
Mechanical circulatory support offers an alternative therapeutic treatment for patients with dysfunctional single ventricle physiology. An intravascular axial flow pump is being developed as a cavopulmonary assist device for these patients. This study details the development of a new rotating impeller geometry. We examined the performance of 8 impeller geometries with blade stagger or twist angles varying from 100° to 800° using computational methods. A refined range of blade twist angles between 300° and 400° was then identified, and 4 additional geometries were evaluated. Generally, the impeller designs produced 4-26mmHg for flow rates of 1-4L/min for 6000-8000 RPM. A data regression analysis was completed and found the impeller with 400° of blade twist to be the superior performer. A hydraulic test was conducted on a prototype of the 400° impeller, which generated measurable pressure rises of 7-28mmHg for flow rates of 1-4L/min at 6000-8000 RPM. The findings of the numerical model and experiment were in reasonable agreement within approximately 20%. These results support the continued development of an axial-flow, mechanical cavopulmonary assist device as a new clinical therapeutic option for Fontan patients.
No takes yet. Share an insight, caveat, or question.
Chopski et al. (2016) studied Dysfunctional single ventricle physiology / Fontan patients. Intravascular axial flow pump impeller designs was evaluated on Pressure rise for given flow rates and RPM. An impeller design with 400° of blade twist generated measurable pressure rises of 7-28 mmHg for flow rates of 1-4 L/min at 6000-8000 RPM in hydraulic testing.
Synapse has enriched 10 related papers. Consider them for comparative context: