Answers are grounded in this paper and cite their sources.
This study examined used, discarded rotary nickel-titanium instruments obtained from 14 endodontists in four countries, and identified factors that may influence defects produced during clinical use. A total of 7,159 instruments were examined for the presence of defects. Unwinding occurred in 12% of instruments and fractures in 5% (1.5% torsional, 3.5% flexural). The defect rates varied significantly among endodontists. Instrument design factors also influenced defect rate, but to a lesser extent. The mean number of uses of instruments with and without defects was 3.3 +/- 1.8 (range: 1-10), and 4.5 +/- 2.0 (range: 1-16), respectively. The most important influence on defect rates was the operator, which may be related to clinical skill or a conscious decision to use instruments a specified number of times or until defects were evident.
No takes yet. Share an insight, caveat, or question.
Parashos et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: