Key result
Retained transvenous pacemakers alongside leadless devices risk emergency explant from end-of-service compulsory pacing.
Case Report (n=1)
No
When implanting a leadless pacemaker instead of replacing a depleted transvenous pacemaker battery, clinicians must account for the remaining device's non-programmable end-of-service behavior, which can trigger compulsory pacing with unstable sensing and necessitate emergency explantation.
Answers are grounded in this paper and cite their sources.
May require preemptive explantation of depleted devices; case report leaves open optimal hybrid pacing strategies.
Micra (Medtronic, Minneapolis, MN) is a leadless pacemaker implanted percutaneously through a transcatheter system1,2 and is an alternative form of ventricular pacing that may avoid the complications of traditional transvenous pacing systems.3 Compared with the traditional transvenous systems, Micra reportedly reduces the incidence of major complications by approximately 63%.4 Therefore, a leadless pacemaker is preferred for older patients with comorbidities who are at significantly increased risk of transvenous pacing systems complications.
No takes yet. Share an insight, caveat, or question.
Ogano et al. (2022) conducted a case report in Transvenous pacemaker battery depletion with history of sick sinus syndrome and persistent atrial fibrillation (n=1). Leadless pacemaker (Micra) implantation without explanting the remaining transvenous pacemaker was evaluated on Compulsory ventricular pacing by the remaining transvenous pacemaker due to end-of-service (EOS) program initiation requiring emergency removal. Leadless pacemaker implantation without explanting an old transvenous pacemaker resulted in compulsory ventricular pacing at 67.5 bpm at 12 months due to end-of-service activation.
Synapse has enriched 10 related papers. Consider them for comparative context: