A liquid chromatographic/tandem mass spectrometric method is described for the determination of phencyclidine (PCP) in small volumes of rat serum (e.g. 50 microl). Samples were extracted using a mixed-mode strong cation-exchange column and then separated isocratically using a narrow-bore (2.1 mm i.d.) 3 microm Hypersil phenyl column and a mobile phase consisting of an ammonium formate buffer (pH 2.7) with 60% (v/v) methanol. Detection was accomplished using positive ion electrospray ionization in the multiple reaction monitoring mode. Mass spectra were obtained and peaks were observed at an m/z (% abundance) of 244 (100), 159 (25), and 86 (89). Tandem mass spectra were also obtained from the m/z 244 precursor ion with peaks observed at m/z 159 (100), 86 (96), and 91 (11). Optimum serum PCP sensitivity and precision were obtained at a transition of m/z 244 --> 159. Matrix-associated ion suppression did not significantly affect the accuracy (100-112%) or precision (CV < or =8%) of the assay. The lower limit of quantitation was 1 ng ml(-1) in 50 microl of serum. The method was used to study the serum pharmacokinetics of PCP in rats after an intravenous bolus dose of PCP.