Does the Rhine-Tec specular microscope agree with the Topcon SP2000P in measuring corneal endothelial cell density?
Does the Rhine-Tec specular microscope agree with the Topcon SP2000P in measuring corneal endothelial cell density?
The Rhine-Tec specular microscope requires manual correction (touched-up mode) to achieve acceptable agreement with the Topcon SP2000P for measuring corneal endothelial cell density.
Rhine-Tec automated endothelial counts should not guide decisions without manual correction; cross-sectional data leaves open device interchangeability in routine practice.
Endothelial cell density (ECD) assessment with the noncontact Topcon SP2000P specular microscope is known to be reliable when the automated mode is followed by manual corrections (touched-up mode). We compared its agreement with Rhine-Tec, a new noncontact specular microscope, in 270 eyes of 160 patients, by comparing the ECD measured in the automated and touched-up modes with that of Topcon touched-up. Good agreement existed between either touched-up modes with a mean difference of only 2 cells/mm2 95% CI (−27; 23) whereas agreement with the Rhine-Tec automated mode was poor with an overestimation by a mean of 226 cells/mm2 95% CI (172; 281). Background The Topcon SP2000P non-contact specular microscope (Topcon, Tokyo, Japan) is widely used to measure corneal endothelial cell density (ECD). It uses a cell contour recognition algorithm based on contrast differences, and ECD derived from an automated delineation of cell boundaries with manual correction of inaccurately drawn cells (“touched-up” mode) has been validated.1,2 A new commercially available non-contact specular microscope Rhine-Tec (Rhine-Tec, Krefeld, Germany) determines ECD by a cell-centre method ...
No takes yet. Share an insight, caveat, or question.
Thuret et al. (2007) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: