Why the study?
Does inert gas rebreathing provide comparable non-invasive cardiac output measurements to cardiac magnetic resonance imaging in patients with atrial fibrillation?
Does inert gas rebreathing provide comparable non-invasive cardiac output measurements to cardiac magnetic resonance imaging in patients with atrial fibrillation?
Non-invasive cardiac output measurement using inert gas rebreathing shows good agreement with cardiac magnetic resonance imaging in patients with atrial fibrillation, except in hyperdynamic circulatory conditions.
Answers are grounded in this paper and cite their sources.
Supports inert gas rebreathing for CO assessment in AF; leaves open prospective validation before clinical use.
OBJECTIVES: Atrial fibrillation (AF) is one of the most frequent heart rhythm disorders. It potentially influences cardiac function and its measurement. Cardiac magnetic resonance imaging (CMR) has become the new gold standard for non-invasive assessment of cardiac output (CO). A novel inert gas rebreathing (IGR) device based on the Fick Principle also proved promising in patients in sinus rhythm (SR). The aim of our study was to compare the agreement of non-invasive CO measurements between CMR and IGR in AF patients. METHODS: A total of 68 patients, 34 with AF and 34 pair-matched controls in SR, were included. RESULTS: Bland-Altman analysis showed good agreement between both methods, with an average deviation of 0.2 +/-1.2 l/min in the AF group versus 0.3 +/-1.0 l/min in the SR group (p = 0.77). IGR demonstrated good agreement for CO between 2.0 and 5.4 l/min. However, in hyperdynamic circulatory conditions (CO >5.5 l/min), the increasing disagreement of IGR and CMR measurements reached statistical significance. CONCLUSIONS: Non-invasive CO measurements using CMR and IGR are feasible in patients suffering from AF. Good agreement was found between the two methods in an unselected cohort. Hyperdynamic circulatory conditions can lead to significant measurement differences which, however, do not affect the reproducibility of IGR.
No takes yet. Share an insight, caveat, or question.
Saur et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: