Answers are grounded in this paper and cite their sources.
A method has been developed to determine the boiling point distribution of sulfur compounds in light cycle oils (LCO'S). The method chosen for this analysis was GC with a flame photometric detector (FPD) and pyrolyzer. Tests were carried out to evaluate the recovery efficiency, repeatability, and accuracy of the method. Repeatabilities within 2% were obtained. The recovery of benzothiophenes and dibenzothiophenes was close to 100%; this was important because these are the major sulfur components in LCO's. No hydrocarbon or solvent interferences were observed with the use of the pyrolyzer, even for a 95% solvent level. Comparison with results from other techniques showed that the method accurately determined the levels of sulfur compounds in the LCO boiling point range.
No takes yet. Share an insight, caveat, or question.
Sung et al. (1994) studied this question.