This paper introduces the design of an ultra fast x-ray tomography scanner based on electron beam technology. The scanner has been developed for two-phase flow studies where frame rates of 1 kHz and higher are required. Its functional principle is similar to that of the electron beam x-ray CT scanners used in cardiac imaging. Thus, the scanner comprises an electron beam generator with a fast beam deflection unit, a semicircular x-ray production target made of tungsten alloy and a circular x-ray detector consisting of 240 CZT elements with 1.5 mm × 1.5 mm × 1.5 mm size each. The design is optimized with respect to ultra fast imaging of smaller flow vessels, such as pipes or laboratory-scale chemical reactors. In that way, the scanner is capable of scanning flow cross-sections at a speed of a few thousand frames per second which is sufficient to capture flows of a few meters per second velocity.