Key result
CanGaroo ECM envelopes were associated with device erosion in 43% of mice compared to 0% with TYRX envelopes, and resulted in thicker capsules at 12 weeks.
Why the study?
Bioabsorbable materials have been developed to help prevent CIED pocket-related complications, but direct comparisons between these adjunctive devices were lacking.
Does the use of CanGaroo ECM or TYRX envelopes prevent device erosion and infection compared to no envelope in a mouse model of CIED implantation?
Does the use of CanGaroo ECM or TYRX envelopes prevent device erosion and infection compared to no envelope in a mouse model of CIED implantation?
In a mouse model, CanGaroo ECM envelopes were associated with a higher rate of device erosion and thicker capsule formation compared to TYRX envelopes or no envelope.
May warrant caution with CanGaroo envelopes in CIED implants; leaves open comparative human efficacy.
INTRODUCTION: Cardiac implantable electronic device (CIED) pocket related problems such as infection, hematoma, and device erosion cause significant morbidity and the clinical consequences are substantial. Bioabsorbable materials have been developed to assist in the prevention of these complications but there has not been any direct comparison of these adjunctive devices to reduce these complications. We sought to directly compare the TYRX absorbable antibacterial and CanGaroo extracellular matrix (ECM) envelopes in an animal model susceptible to these specific CIED-related complications (i.e., skin erosion and infection). METHODS AND RESULTS: Sixteen mice undergoing implantation with biopotential transmitters were divided into three groups (no envelope = 4, TYRX = 5, and CanGaroo = 7) and monitored for device-related complications. Following 12 weeks of implantation, gross and histological analysis of the remaining capsules was performed. Three animals in the CanGaroo group (43%) had device erosion compared to none in the TYRX group. The remaining capsules excised at 12 weeks were qualitatively thicker following CanGaroo compared to TYRX and no envelope and histological evaluation demonstrated increased connective tissue with CanGaroo. CONCLUSION: CanGaroo ECM envelopes did not reduce the incidence of device erosion and were associated with qualitatively thicker capsules and connective tissue staining at 12 weeks compared to no envelope or TYRX. Further studies regarding the use of these envelopes to prevent device erosion and their subsequent impact on capsule formation are warranted.
No takes yet. Share an insight, caveat, or question.
Ip et al. (2020) studied CIED-related complications (n=16). CanGaroo extracellular matrix (ECM) envelope vs. TYRX absorbable antibacterial envelope and no envelope was evaluated on Device erosion. CanGaroo ECM envelopes were associated with device erosion in 43% of mice compared to 0% with TYRX envelopes, and resulted in thicker capsules at 12 weeks.