Answers are grounded in this paper and cite their sources.
We investigated the thermal properties of a new bone cement, Boneloc, during total hip replacement, using a thermocouple connected to a digital thermometer. In 10 of 11 cases the maximum temperature was below 43 degrees C at the bone-cement interface in the acetabulum. This is a considerable reduction in the polymerization temperature in comparison with earlier studies of conventional bone cement. We found no difference regarding the setting time of the cement.
No takes yet. Share an insight, caveat, or question.
Wykman et al. (1995) studied this question.