Key result
FSE T2 FS, STIR, and IDEAL achieved more accurate fat suppression fractions than chemical presaturation methods on a water-fat phantom at 1.5 T.
In a phantom model at 1.5 T, IDEAL, FSE T2 FS, and STIR provided the most accurate fat suppression quantification, with IDEAL yielding the highest SNR and CNR.
Phantom comparisons of FS methods may inform protocol selection; leaves open in vivo validation before clinical use.
Nowadays, magnetic resonance imaging (MRI) has been widely applied for diagnosis of soft-tissue diseases. Most clinical MRI protocols use fat suppression (FS) methods to suppress fat signal, reduce chemical shift artifacts, and increase conspicuity of lesions. To understand the advantages, disadvantages, and clinical applications of the most commonly used FS methods is an important issue. The aim of this study was to evaluate FS performance of six FS methods on a fat-water phantom at 1.5[Formula: see text]T. The six MRI methods included iterative decomposition of water and fat with echo asymmetry and least squares estimation (IDEAL), short inversion time inversion recovery (STIR), and four chemical presaturation (Chem Presat) methods. The phantom was composed of homogeneous oil-in-water emulsions with various fat contents ranging from 0 to 100% in increments of 10%. The difference between the suppressed fat fractions (FS fractions) and the true fat fractions of the phantom was used as an index of FS performance. The correlations and levels of agreement (LOAs) between the FS fractions determined using each FS method and the true fat fractions of the phantom were analyzed. From the phantom study, it was found that FSE T2 FS, STIR and IDEAL could achieve more accurate FS fractions than the other three methods. The FS fractions determined using FSE T2 FS, STIR and IDEAL were in a good agreement. On the contrary, T2-weighted spin echo Chem Presat had the most inaccurate quantification of FS fractions among these six FS methods. Both the ranks of signal-to-noise ratios (SNRs) and contrast-to-noise ratios (CNRs) of the phantom were IDEAL [Formula: see text] FSE T2 FS [Formula: see text] STIR. The FS performance of these six FS methods in clinical use needs further study.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2017) studied this question. MRI fat suppression methods (IDEAL, STIR, and four Chem Presat methods) vs. True fat fractions of the phantom was evaluated on Difference between suppressed fat fractions and true fat fractions. FSE T2 FS, STIR, and IDEAL achieved more accurate fat suppression fractions than chemical presaturation methods on a water-fat phantom at 1.5 T.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: