Mechanical circulatory support options for infants and children are very limited in the United States. Existing circulatory support systems have proven successful for short-term pediatric assist, but are not completely successful as a bridge-to-transplant or bridge-to-recovery. To address this substantial need for alternative pediatric mechanical assist, we are developing a novel, magnetically levitated, axial flow pediatric ventricular assist device (PVAD) intended for longer-term ventricular support. Three major numerical design and optimization phases have been completed. A prototype was built based on the latest numerical design (PVAD3) and hydraulically tested in a flow loop. The plastic PVAD prototype delivered 0.5-4 lpm, generating pressure rises of 50-115 mm Hg for operating speeds of 6,000-9,000 rpm. The experimental testing data and the numerical predictions correlated well. The error between these sets of data was found to be generally 7.8% with a maximum deviation of 24% at higher flow rates. The axial fluid forces for the numerical simulations ranged from 0.5 to 1 N and deviated from the experimental results by generally 8.5% with a maximum deviation of 12% at higher flow rates. These hydraulic results demonstrate the excellent performance of the PVAD3 and illustrate the achievement of the design objectives.