BACKGROUND AND OBJECTIVES: Endoscopic applications of the Erbium:YAG laser have been limited due to the lack of a suitable optical fiber delivery system. The purpose of this study was to compare the transmission of Er:YAG laser radiation through germanium oxide trunk fibers with silica and sapphire fiber tips for potential use in contact tissue ablation during endoscopy. STUDY DESIGN/MATERIALS AND METHODS: Er:YAG laser radiation with a wavelength of 2.94 microm, pulse length of 300 microseconds, pulse energies from 5 to 1,360 mJ, coupled into pulse repetition rates of 3-10 Hz, was through 1-m-long germanium oxide fibers with either 1-cm-long, 550-microm-diameter silica or sapphire tips. RESULTS: Transmission through the germanium oxide/sapphire fibers measured 65+/-5% compared with 55+/-4% for the germanium oxide/silica fibers (P<0.05). The damage threshold for the hybrid fibers averaged 309+/- 44 mJ and 126+/-43 mJ, respectively (n = 7 fibers each) (P<0.05). The highest pulse energies transmitted through the fibers were 700 mJ and 220 mJ, respectively. CONCLUSIONS: Improved index-matching of the trunk fiber and fiber tip at 2.94 microm resulted in higher transmission and damage thresholds for the germanium oxide/sapphire fibers. The germanium oxide/sapphire fiber may represent a promising mid-infrared optical fiber delivery system for use in endoscopic applications of the Er:YAG laser requiring a flexible, biocompatible, and robust fiber delivery system for contact tissue ablation.