Clarifies magnetic resonance imaging safety risks for implanted clips, highlighting that only stainless steel implants possess dangerous ferromagnetic properties.
The problem of implanted metals causing tissue damage by movement in patients exposed to MRI fields has produced a confusing welter of erroneous, pseudoscientific publications about magnetics, metals, medical equipment, and tissue compatibility. Quite simply, among the devices made for implantation, only those fabricated of stainless steel have the ferromagnetic properties capable of causing such accidents. The author, who introduced the basic design of the modern aneurysm clip in the late 1960s and then a cobalt nickel alloy as an improvement over steel, while chairing the neurosurgical committee assigned to the task of establishing neurosurgical standards at American Society for Testing and Materials, exposes this flawed information and offers clear guidelines for avoiding trouble.
No takes yet. Share an insight, caveat, or question.
Joseph T. McFadden (2012) studied this question.