Answers are grounded in this paper and cite their sources.
A simple method is described for the approximate measurement of the ion energy distribution in an inductively coupled plasma-mass spectrometer (ICP-MS) with a continuum flow sampling interface. The average ion kinetic energy, kinetic energy spread, and maximum kinetic energy are evaluated from a plot of ion signal as a function of retarding voltage applied to the quadrupole mass analyzer. The effects of plasma operating parameters on ion signals and energies are described. In particular, kinetic energy is a sensitive function of aerosol gas flow rate. This behavior is attributed to a nonthermal, possibly electrical, interaction between the plasma and the sampling interface, which is induced by the presence of the axial channel in the ICP.
No takes yet. Share an insight, caveat, or question.
Olivares et al. (1985) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: