Key result
Transesophageal echocardiography measured a smaller maximum LAA ostium diameter than X-ray angiography during LAAO in NOHD patients.
Why the study?
The intraoperative guiding role and postoperative evaluative function of transesophageal echocardiography during left atrial appendage occlusion in patients with non-organic heart disease needed to be assessed.
Does transesophageal echocardiography (TEE) accurately guide device sizing, monitor anatomical changes, and detect peri-device leaks compared to X-ray angiography during and after left atrial appendage occlusion in patients with non-organic heart disease?
Cohort (n=48)
No
Does transesophageal echocardiography (TEE) accurately guide device sizing, monitor anatomical changes, and detect peri-device leaks compared to X-ray angiography during and after left atrial appendage occlusion in patients with non-organic heart disease?
Answers are grounded in this paper and cite their sources.
In patients with non-organic heart disease undergoing LAAO, TEE yields significantly smaller LAA ostium and depth dimensions than X-ray angiography while effectively guiding transseptal puncture, device sizing, and detection of minor peri-device leaks.
OBJECTIVE: To assess the guiding role of transesophageal echocardiography (TEE) intraoperatively and its evaluative function postoperatively during left atrial appendage occlusion (LAAO) in patients with non-organic heart disease (NOHD). METHODS: In this retrospective observational study, a total of 48 patients with NOHD who underwent LAAO in the Department of Cardiology at The First People's Hospital of Lanzhou City from April 2020 to September 2022 were recruited. TEE findings during and after the procedure, cardiac chamber size, and cardiac function parameters at different surgical stages, postoperative occlusion efficacy, and complications were recorded. The application value of TEE in LAAO for patients with NOHD was evaluated. RESULTS: Comparative analysis TEE-measured the maximum diameter of the LAAO (22.37 ± 3.86 mm) was significantly smaller than X-ray angiographic measurement (23.45 ± 4.22 mm; p < 0.05). One month after radiofrequency ablation, TTE revealed a statistically significant reduction in left atrial diameter (p < 0.05). Four cases (8%) exhibited minor peri-device leak (< 3 mm), and no major complications occurred. CONCLUSION: TEE shows significant application value for monitoring anatomical changes, guiding device sizing, and detecting peri-device leaks during and after LAAO for patients with NOHD.
No takes yet. Share an insight, caveat, or question.
Long et al. (2025) conducted a cohort in Non-organic heart disease (NOHD) undergoing left atrial appendage occlusion (LAAO) (n=48). Transesophageal echocardiography (TEE) intraoperatively during LAAO vs. X-ray angiography (XRA) measurement was evaluated on Maximum diameter of the left atrial appendage (LAA) ostium (mm) (p=0.04). Transesophageal echocardiography measured a smaller maximum LAA ostium diameter than X-ray angiography during LAAO in NOHD patients.