Key result
Intraoperative hypothermia was associated with excessive defibrillation thresholds in two patients with newly implanted ICDs, which normalized during postoperative normothermia.
Why the study?
Does inadvertent intraoperative hypothermia increase defibrillation thresholds in patients undergoing AICD implantation?
Case Report (n=2)
Does inadvertent intraoperative hypothermia increase defibrillation thresholds in patients undergoing AICD implantation?
Inadvertent intraoperative hypothermia during AICD implantation may increase defibrillation thresholds, highlighting the importance of temperature management during device testing.
Answers are grounded in this paper and cite their sources.
May warrant temperature monitoring during ICD implantation; hypothesis-generating for hypothermia effects on thresholds.
We describe two patients with newly implanted automatic implantable cardioverter-defibrillators that had excessive defibrillation thresholds associated with hypothermia at intraoperative defibrillation threshold testing. Normal defibrillation threshold levels were obtained during postoperative non-invasive electrophysiology testing in an electrophysiology laboratory when the patients were normothermic. We hypothesize that inadvertent intraoperative hypothermia during device implantation may increase the defibrillation threshold.
No takes yet. Share an insight, caveat, or question.
Kapetanopoulos et al. (2007) conducted a case report in Newly implanted automatic implantable cardioverter-defibrillators (n=2). Intraoperative hypothermia vs. Normothermia was evaluated on Defibrillation threshold. Intraoperative hypothermia was associated with excessive defibrillation thresholds in two patients with newly implanted ICDs, which normalized during postoperative normothermia.
Synapse has enriched 10 related papers. Consider them for comparative context: