Key result
Automated ECG devices produced high false negative rates, failing to detect QTcF >450 ms in 39.5% of cases and QTcF >500 ms in 55.8% of cases compared to centralized core lab measurements.
Why the study?
Clinical trial sites often lack the expertise and infrastructure to accurately measure cardiac intervals on 12-lead ECGs and rely heavily on automated device-generated results for clinical decisions.
Do automated ECG device measurements accurately detect QTcF and QRS prolongation compared to centralized core lab measurements in clinical oncology trials?
Observational (n=17,475)
Yes
Do automated ECG device measurements accurately detect QTcF and QRS prolongation compared to centralized core lab measurements in clinical oncology trials?
Automated ECG algorithms frequently fail to detect clinically significant QTc and QRS prolongation, highlighting the danger of relying solely on unoverread automated ECG results for safety decisions in clinical trials.
Answers are grounded in this paper and cite their sources.
Automated QTc readings require core lab confirmation in oncology trials; this cross-sectional analysis leaves open prospective device validation.
BACKGROUND: Sites participating in clinical trials may not have the expertise and infrastructure to accurately measure cardiac intervals on 12-lead ECGs and rely heavily on the automated ECG device generated results for clinical decision-making. METHODS: Using a dataset of over 260,000 ECGs collected in clinical oncology studies, we investigated the mean difference and the rate of false negative results between the digital ECG machine QTc and QRS measurements compared to those obtained by a centralized ECG core lab. RESULTS: The mean differences between the core lab and the automated algorithm QTcF and QRS measurements were + 1.8 ± 16.0 ms and - 1.0 ± 8.8 ms, respectively. Among the ECGs with a centralized QTcF value > 450 or > 470 ms, 39.5% and 47.8% respectively had a device reported QTcF value ≤ 450 ms or ≤ 470 ms. Among the ECGs with a centrally measured QTcF > 500 ms, 55.8% had a device reported value ≤ 500 ms. Automated QTcF measurements failed to detect a QTcF increase > 60 ms for 53.9% of the ECGs identified by the core lab. Automated measurements also failed to detect QRS prolongation, though to a lesser extent than failures to detect QTc prolongation. Among the ECGs with a centrally measured QRS > 110 or 120 ms, 7.9% and 7.3% respectively had a device reported QRS value ≤ 110 ms or ≤ 120 ms. CONCLUSION: Relying on automated measurements from ECG devices for patient inclusion and treatment (dis)continuation decisions poses a potential risk to patients participating in oncology studies.
No takes yet. Share an insight, caveat, or question.
Dekie et al. (2022) conducted an observational in Oncology (n=17,475). Automated ECG device measurements vs. Centralized ECG core lab measurements was evaluated on Mean difference in QTcF between core lab and automated algorithm (MD 1.8). Automated ECG devices produced high false negative rates, failing to detect QTcF >450 ms in 39.5% of cases and QTcF >500 ms in 55.8% of cases compared to centralized core lab measurements.