Key result
Inadvertent catheter-induced right bundle branch block prevented the initiation of bundle branch reentry ventricular tachycardia, highlighting a risk for misinterpreting drug efficacy.
Case Report (n=1)
Inadvertent catheter-induced right bundle branch block can temporarily prevent the induction of bundle branch reentrant ventricular tachycardia, potentially leading to misinterpretation of antiarrhythmic drug efficacy.
Answers are grounded in this paper and cite their sources.
Catheter-induced RBBB may confound BBRVT induction; case reports leave open reliable assessment of antiarrhythmic drug efficacy.
This article describes the inadvertent, catheter-induced induction of right bundle branch block resulting not only in transient complete infra-His heart block but also in temporary interruption of the macroreentry circuit of ventricular tachycardia. A patient with preexistent left bundle branch block and spontaneous ventricular tachycardia based upon the bundle branch reentry mechanism underwent electrophysiological testing for the evaluation of sotalol drug efficacy. In search of an optimal His-bundle recording, the manipulation of a 6 Fr quadripolar catheter caused a right bundle branch block, thus advancing the preexistent left bundle branch block to complete heart block. Retrograde ventriculoatrial conduction remained unaffected. The macroreentrant tachycardia with left bundle branch block configuration was no longer inducible. While the patient continued on unchanged sotalol medication (320 mg/d) he required temporary pacing for 16 hours until the block subsided. A subsequent induction attempt demonstrated initiation of the tachycardia. Finally, guided by invasive testing, the patient successfully received amiodarone therapy (300 mg/d). The patient completed an uneventful follow up of 27 months. No progression of conduction delay was observed. This case suggests that the inadvertent induction of right bundle branch block prevents the initiation of ventricular tachycardias relying on bundle branch reentry. Therefore, missed diagnosis or misinterpretation of antiarrhythmic drug efficacy might occur if there is no electrophysiological reevaluation after right bundle branch recovery.
No takes yet. Share an insight, caveat, or question.
Gössinger et al. (1989) conducted a case report in Preexistent left bundle branch block and recurrent macroreentrant ventricular tachycardia (n=1). Inadvertent catheter-induced right bundle branch block was evaluated on Induction of ventricular tachycardia. Inadvertent catheter-induced right bundle branch block prevented the initiation of bundle branch reentry ventricular tachycardia, highlighting a risk for misinterpreting drug efficacy.