Answers are grounded in this paper and cite their sources.
In this resistivity measurement, a pair of semicircular measuring electrodes is connected to a series or parallel resonance circuit composed of an inductance L and a variable capacitance C. A sample is coupled capacitively with the electrode with no direct contact. The authors' method is similar to the Q meter method in principle, but uses a comparative measurement. Some analyses of this method and the experimental results are reported: (1) the measurable range of resistivity is limited by an electrode structure and a frequency, (2) the resistivity measurement over a range as wide as 0.3 to 1500 Ω-cm can be carried out by using both the series and the parallel resonance circuits, (3) for a thin sample below 3 mm in thickness, the resistivity can be measured by using a calibration, (4) the influence of different surface treatments is not seen if the surface remains flat.
No takes yet. Share an insight, caveat, or question.
Miyamoto et al. (1967) studied this question.