This study evaluated the accuracy of a two-ball bearing radiographic marker on assessing the true magnification of post-operative total hip arthroplasty (THA) films. The effects of body mass index (BMI) and height on the measured error rates were also examined. We retrospectively reviewed 211 consecutive THA films. The implanted hemispherical acetabular component size was used to calculate the actual magnification. The average film magnification was 118.7%. There was a mean measured error rate of 1.27% (SD = 3.32) for this marker. No significant difference was found between measurements on consecutive post-operative films (n=52). BMI and height did not demonstrate any effect on error rate. These results support the use of this two-ball bearing marker as a reliable instrument for assessing x-ray magnification in pre-operative templating.