The mobilities of mass-identified U+ and Hg+ ions in helium have been determined in a drift tube-mass spectrometer apparatus. For uranium ions a reduced mobility value μ0=(16.0± 0.5) cm2/V·sec is obtained at 305°K and a standard gas density of 2.69× 1019cm−3. The mobility of mercury ions is μ0=(19.4± 0.5) cm2/V·sec at 292°K, in agreement with two previous determinations. The effect of fast ion injection in drift mobility measurements is discussed, and a new technique to circumvent these problems is described. The results are compared with existing theories of ion mobilities.
No takes yet. Share an insight, caveat, or question.
Johnsen et al. (1972) studied this question.