Key result
A 1D convolutional neural network accurately estimated amplitude spectrum area during uninterrupted cardiopulmonary resuscitation, achieving a mean absolute error of 1.951 mVHz on real-life data.
Why the study?
Accurate amplitude spectrum area (AMSA) can only be calculated during cardiopulmonary resuscitation (CPR) pauses due to artifacts from chest compressions.
Can a 1D convolutional neural network accurately estimate amplitude spectrum area (AMSA) during uninterrupted chest compressions in patients with ventricular fibrillation?
Can a 1D convolutional neural network accurately estimate amplitude spectrum area (AMSA) during uninterrupted chest compressions in patients with ventricular fibrillation?
Answers are grounded in this paper and cite their sources.
A novel 1D CNN algorithm can accurately estimate AMSA during uninterrupted CPR, potentially guiding individualized resuscitation without requiring CPR pauses.
Introduction: Amplitude spectrum area (AMSA) is a well-established measure than can predict defibrillation outcome and guiding individualized resuscitation of ventricular fibrillation (VF) patients. However, accurate AMSA can only be calculated during cardiopulmonary resuscitation (CPR) pause due to artifacts produced by chest compression (CC). In this study, we developed a real-time AMSA estimation algorithm using a convolutional neural network (CNN). Methods: Data were collected from 698 patients, and the AMSA calculated from the uncorrupted signals served as the true value for both uncorrupted and the adjacent corrupted signals. An architecture consisting of a 6-layer 1D CNN and 3 fully connected layers was developed for AMSA estimation. A 5-fold cross-validation procedure was used to train, validate and optimize the algorithm. An independent testing set comprised of simulated data, real-life CC corrupted data, and preshock data was used to evaluate the performance. Results: The mean absolute error, root mean square error, percentage root mean square difference and correlation coefficient were 2.182/1.951 mVHz, 2.957/2.574 mVHz, 22.887/28.649% and 0.804/0.888 for simulated and real-life testing data, respectively. The area under the receiver operating characteristic curve regarding predicting defibrillation success was 0.835, which was comparable to that of 0.849 using the true value of the AMSA. Conclusions: AMSA can be accurately estimated during uninterrupted CPR using the proposed method.
No takes yet. Share an insight, caveat, or question.
Zuo et al. (2023) studied Out-of-hospital cardiac arrest with ventricular fibrillation (n=728). 1D convolutional neural network (CNN) algorithm vs. Fast Fourier transformation (FFT) and adaptive filtering (ADF) methods was evaluated on Mean absolute error (MAE) of AMSA estimation on real-life testing data (p=<0.05). A 1D convolutional neural network accurately estimated amplitude spectrum area during uninterrupted cardiopulmonary resuscitation, achieving a mean absolute error of 1.951 mVHz on real-life data.
Synapse has enriched 10 related papers. Consider them for comparative context: