A novel 14F double lumen catheter demonstrated adequate hemodynamic and oxygen transport characteristics in preclinical testing, supporting its potential use in neonatal venovenous ECMO.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for neonatal VV-ECMO; leaves open translation from canine models to clinical use.
The authors designed and tested a 14F outside diameter thin-walled double lumen catheter (DLC) for neonatal venovenous (VV) extracorporeal membrane oxygenation (ECMO). In vitro tests with water and dye solution showed capacity of the drainage lumen was 1,096 ml/min at 100cm siphon, and pressure drop across the perfusion lumen was 300 mmHg at 500 ml/min flow. Recirculation at 500 ml/min flow ranged from 5 to 29%, depending upon simulated cardiac output. The highest serum hemoglobin during 12 hour 400 ml/min flow VV bypass in five dogs was 49 mg/dl. Typical oxygen transport in four dogs was 25 cc/min at 400 ml/min flow. This catheter is well suited for clinical VV ECMO in neonates.