This is the authors' abstract. We don't add key points for this paper.
Typical ionization techniques used for mass spectrometry (MS) analysis face challenges when trying to analyze organic species in a high-salt environment. Here, we present results using a recently developed ionization source, liquid sampling-atmospheric pressure glow discharge (LS-APGD), for marine-relevant salt-containing organic samples. Using two representative sample types, a triglyceride mixture and dissolved organic matter, this method is compared to traditional electrospray ionization (ESI) under saline and neat conditions. LS-APGD produced equal or higher (15%+) ion intensities than those of ESI for both salt-containing and neat samples, although important differences linked with adduct formation in high-salt conditions explain the molecular species observed. For all sample types, LS-APGD observed a higher diversity of molecules under optimized settings (0.25 mm electrode spacing at 20 mA) compared to traditional ESI. Furthermore, because the LS-APGD source ionizes molecular species in a ∼1 mm 3 volume plasma using a low-power source, there is the potential for this method to be applied in field studies, eliminating desalting procedures, which can be time-consuming and nonideal for low-concentration species.