Answers are grounded in this paper and cite their sources.
Compositional analysis of interfaces in semiconductor materials grown with the most modern equipment requires a substantial improvement of the depth resolution of secondary ion mass spectrometry (SIMS). The lowering of the impact energy to improve depth resolution is limited on most magnetic-sector instruments to ∼1.5 keV. In this work it is shown that in the VG IX70S magnetic-sector instrument a reduction of the impact energy to 600 eV is possible. The reduction is achieved by use of a deceleration electrode in the primary beam line, allowing for independent variation of the energy and the incidence angle θ (50°<θ<80°). The best depth resolution obtained—for a shallow Ge δ layer with a 600 eV θ=80° O2+ beam—was 1.6 nm full width at half-maximum and an exponential decay length of 0.65 nm; about three times better than achievable on any other magnetic-sector SIMS instrument. In addition, a very shallow (∼3–4 nm) Ge δ layer has been analyzed by high-resolution Rutherford backscattering and SIMS. There is good agreement between both techniques.
No takes yet. Share an insight, caveat, or question.
Alkemade et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: