Diagnostic validation study demonstrates rapid HIV-1 antibody detection in whole blood samples, suggesting great potential for simple qualitative screening in medical and veterinary care.
An antibody detection procedure based on agglutination of autologous red cells has been developed for samples of whole blood. A nonagglutinating monoclonal antibody to human red blood cells conjugated to a synthetic peptide antigen (in this case residues 579 to 601 of the HIV-1 envelope precursor, Arg-Ile-Leu-Ala-Val-Glu-Arg-Tyr-Leu-Lys-Asp-Gln-Gln-Leu-Leu-Gly-Ile-Trp- Gly-Cys - Ser-Gly-Lys) permitted the detection of antibodies to the human immunodeficiency virus type 1 (HIV-1) in 10 microliters of whole blood within 2 minutes. Agglutination was specifically inhibited by addition of synthetic peptide antigen but not by unrelated peptides. The frequency of false positive results was 0.1% with HIV-1 seronegative blood donors (n = 874). The false negative results were approximately 1% (n = 81). The autologous red cell agglutination test is potentially suitable for simple, rapid, qualitative screening for antibodies to a variety of antigens of medical and veterinary diagnostic significance.