Answers are grounded in this paper and cite their sources.
A method is described for improving column stability and reproducibility for analysis of vitamin E vitamers on normal‐phase high‐performance liquid chromatography columns with isooctane/ethyl acetate/acetic acid/2,2‐dimethoxypropane (98.5:0.9:0.85:0.1) mobile phase. Acetic acid in the mobile phase reduced retention times of vitamin E vitamer and increased column stability. A small concentration of 2,2‐dimethoxypropane, which reacts with water to form acetone and methanol, reduced the need for column regeneration and stabilized retention times.
No takes yet. Share an insight, caveat, or question.
J. Samuel Godber (1993) studied this question.