Key result
Noninvasive scalp DC-electroencephalography successfully detected cortical spreading depolarizations with a median Coefficient-of-Determination of 0.98 compared to invasive ECoG.
Why the study?
Cortical spreading depolarization (SD) is increasingly recognized as a major contributor to secondary brain injury, and noninvasive SD monitoring would enable the institution of SD-based therapeutics.
Can high-resolution direct current-electroencephalography noninvasively detect cortical spreading depolarization compared to invasive electrocorticography in brain injury patients?
Observational
Can high-resolution direct current-electroencephalography noninvasively detect cortical spreading depolarization compared to invasive electrocorticography in brain injury patients?
Noninvasive detection of cortical spreading depolarization using scalp DC-potential signals with a high spatial resolution EEG array is possible, though current prototypes are limited by high artifact burden.
Answers are grounded in this paper and cite their sources.
May enable noninvasive SD monitoring in brain injury; leaves open clinical utility pending prospective validation.
Background/Objective Cortical spreading depolarization (SD) is increasingly recognized as a major contributor to secondary brain injury. Noninvasive SD monitoring would enable the institution of SD-based therapeutics. Our primary objective is to establish proof-of-concept validation that scalp DC-potentials can provide noninvasive SD detection by comparing scalp direct-current (DC)-shifts from a high-density electrode array to SDs detected by gold-standard electrocorticography (ECoG). Our secondary objective is to assess usability and artifact tolerance. Methods An 83x58 mm thermoplastic elastomer array with 29 6-mm diameter Ag/AgCl 1-cm spaced electrodes, the CerebroPatch™ Proof-of-Concept Prototype, was adhesively placed on the forehead with an intervening electrode gel interface to record DC-electroencephalography in normal volunteers and severe acute brain injury patients in the neuro-intensive care unit some with and some without invasive ECoG electrodes. The scalp and ECoG voltages were collected by a Moberg® Advanced ICU Amplifier. Artifacts were visually identified and usability issues were recorded. SD was scored on ECoG based on DC-shifts with associated high-frequency suppression and propagation. A six-parameter Gaussian plus quadratic baseline model was used to estimate ECoG and scalp electrode time-courses and scalp-voltage heat-map movies. The similarity of the noninvasive scalp and invasive ECoG DC-shift time-courses was compared via the Gaussian fit parameters and confirmed if the Coefficient-of-Determination was >0.80. Results Usability and artifact issues obscured most scalp Prototype device data of the 140 ECoG-coded SDs during 11 days in one sub-arachnoid hemorrhage patient. Twenty-six of these DC-shifts were in readable, artifact-free portions of scalp recordings and 24 of these had a >0.80 Coefficient-of-Determination (0.98[0.02], median[IQR]) between invasive ECoG and noninvasive Prototype device DC-shifts. Reconstructed heat-map movies of the scalp DC-potentials showed a 5-cm extent, -460 µV peak region that persisted for ∼70 sec. These data suggest that these scalp DC-shifts (peak -457±69 µV [mean±StD], full-width-half maximum 70.9±5.92 sec, area 18.7±2.76 cm 2 ) depicted in the heat-map movies represent noninvasively detected SDs. Conclusions These results using 26 SDs as the observational units suggest that noninvasive SD detection is possible using scalp DC-potential signals with a high spatial resolution EEG array. Although the high artifact burden data and low usability records were limiting, negative results, they serve as an important entrepreneurial recipe that provides suggestions for a future, re-designed device that would reduce artifacts and improve usability for DC-EEG SD detection needed to enable multi-modal monitoring for secondary brain injury.
No takes yet. Share an insight, caveat, or question.
Brown et al. (2024) conducted an observational in Severe acute brain injury. High resolution direct current-electroencephalography (CerebroPatch Prototype) vs. Invasive electrocorticography (ECoG) was evaluated on Similarity of noninvasive scalp and invasive ECoG DC-shift time-courses (Median Coefficient-of-Determination 0.98). Noninvasive scalp DC-electroencephalography successfully detected cortical spreading depolarizations with a median Coefficient-of-Determination of 0.98 compared to invasive ECoG.