In a bovine cervical spine model, the ultimate and fatigue strengths as well as relative magnetic resonance imaging artifact produced by titanium, cobalt chrome, and stainless-steel wires in various gauges were assessed. Single-cycle and fatigue strength of wire constructs were measured. Although larger wires generally had greater static strength, fatigue strength was mixed. Sixteen-gauge titanium, and all stainless-steel models (22-gauge braided, 18-gauge, and Songer cable) withstood 10,000 cycles without failure, whereas all other constructs rarely could withstand a similar 10,000 cycles. Magnetic resonance imaging was performed on calf cervical spines instrumented with the various materials. Titanium exhibited the least artifact, stainless-steel showed the greatest artifact, and cobalt chrome an intermediate amount. Although titanium wire produces the least amount of magnetic resonance imaging artifact, it remains a poor choice for implant fixation because its notch sensitivity reduces its fatigue resistance compared with stainless steel, which remains the more dependable choice.