Key result
Lead extraction of Accufix atrial J-wire pacemaker leads was associated with 13 deaths in 4,076 procedures (0.32%), while 6 deaths were attributed to spontaneous J-wire injury in unextracted leads.
Why the study?
Does lead extraction compared to conservative management affect the rate of injuries and fatalities in patients with Accufix atrial J-wire permanent pacemaker leads?
Observational (n=30,357)
Yes
Does lead extraction compared to conservative management affect the rate of injuries and fatalities in patients with Accufix atrial J-wire permanent pacemaker leads?
Absolute Event Rate: 0.32% vs 0.03%
Awareness of extraction-related complications compared to spontaneous fracture risks highlights the need to carefully balance the risks of patient injury when managing Accufix J-wire pacemaker leads.
Extraction mortality exceeded spontaneous injury deaths; leaves open optimal individualized management of Accufix J-wire leads.
UNLABELLED: To make recommendations for management of potentially fatal failure of the Accufix series of atrial J-wire permanent pacemaker leads, we closely monitored the number of injuries and fatalities resulting either from spontaneous fracture of the J-wire or from attempts to extract the lead. In a population of 30,357 patients, 2,298 patients are enrolled in a prospective follow-up Multicenter Study, the remainder are patients with known clinical status from voluntary reporting, and 2,992 patients died following implant. In the remaining 27,365 patients, 6 deaths have been attributed to J-wire related injury (J-inj) while 13 were complications (E-inj) associated with 4,076 lead extraction procedures (3,974 intravascular (intra)/102 primary thoracotomy (PT). The date of occurrences were from 1994 to November 1997. CONCLUSIONS: (1) Since lead extractions were not conducted in a controlled study, it is not known whether the deaths associated with lead extraction is in excess of what would have occurred if these leads had not been removed in this specific subset. (2) Awareness of the procedure related complication rate appears to have moderated the rate of lead extraction and may ultimately lead to management that reasonably balances the risks of patient injury.
No takes yet. Share an insight, caveat, or question.
Kawanishi et al. (1998) conducted an observational in Accufix series of atrial J-wire permanent pacemaker leads (n=30,357). Lead extraction vs. Conservative management (spontaneous fracture monitoring) was evaluated on Deaths attributed to J-wire related injury or complications associated with lead extraction procedures. Lead extraction of Accufix atrial J-wire pacemaker leads was associated with 13 deaths in 4,076 procedures (0.32%), while 6 deaths were attributed to spontaneous J-wire injury in unextracted leads.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: