Key result
Snare removal of the protruding retention wire via a femoral approach successfully removed the wire in 66.7% (6 of 9) of patients with Class III Accufix leads, with no complications.
Why the study?
Does snare removal via a femoral approach safely remove protruding retention wires in patients with Class III Accufix atrial 'J' leads?
Observational (n=9)
Does snare removal via a femoral approach safely remove protruding retention wires in patients with Class III Accufix atrial 'J' leads?
Snare removal of protruding retention wires via a femoral approach is a feasible and safe option for managing Class III Accufix leads without requiring full lead extraction.
May offer a less invasive option for Class III Accufix leads; hypothesis-generating and requires prospective validation before wider adoption.
The Telectronics Accufix Atrial "J" pacing lead poses a mechanical risk to patients of retention wire fracture and protrusion. Standard lead extraction techniques include percutaneous approaches, which are associated with significant risk of morbidity and mortality, and open procedures, which necessitate thoracotomy. In nine patients referred with Class III retention wire fractures, attempts were made to snare the protruding retention wire from a femoral approach using snare devices and bioptomes. In six cases, the retention wire was successfully removed, leaving the lead body in place. In four patients with lead function that was able to be evaluated, the atrial lead remained functional after the procedure. There were no complications. Snare removal of the protruding retention wire via a femoral approach should be considered as an option in the management of patients with Class III Accufix leads.
No takes yet. Share an insight, caveat, or question.
Ganz et al. (1996) conducted an observational in Class III retention wire fractures in Telectronics Accufix Atrial "J" pacing leads (n=9). Snare removal of the protruding retention wire via a femoral approach was evaluated on Successful removal of the retention wire leaving the lead body in place. Snare removal of the protruding retention wire via a femoral approach successfully removed the wire in 66.7% (6 of 9) of patients with Class III Accufix leads, with no complications.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: