Answers are grounded in this paper and cite their sources.
Three hundred and eighty computer-assisted total knee arthroplasty cases were reviewed for accuracy of mechanical alignment. The 331 patients in the first set, designated Group A, showed a consistent valgus error of 1° from neutral alignment. It was hypothesized that a manual 1° varus correction during femur resection would yield a significantly greater level of accuracy in the second set of 49 patients, designated Group B. A mechanical alignment of ±3° was achieved in 91% of the uncalibrated Group A patients, which was significantly lower (p = 0.035) than the rate of 98% achieved in the calibrated Group B. Further statistical analysis of the data showed the time expenditure was not significantly changed once a new target value was recalibrated. By quantifying mean errors of measures at an early timeframe, enhanced accuracy in CAS can be achieved.
No takes yet. Share an insight, caveat, or question.
Lionberger et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: