Answers are grounded in this paper and cite their sources.
Percutaneous femoral distal epiphysiodesis using transphyseal cannulated screws is commonly used for correcting anisomelia and angular deformity in adolescent patients. Method: We report a case of previously unreported complication using cannulated screws: The thread being damaged, the screw failed to block growth in the physis showed by the releasing of a metal filament in the bone marrow after the screw's removal. Results: The patient showed a limb length discrepancy necessitating a surgical correction. The screw thread was damaged by the other device releasing a metal filament lodged in the bone marrow. The metal filament was released during the first operative gesture for Patient A, whereas it occurred on the removal of the hardwire on the other example.
No takes yet. Share an insight, caveat, or question.
Edouard et al. (2020) studied this question.