Why the study?
Interactions among components of tactical aircraft life support systems may generate unexpected breathing loads on aircrew.
Does low regulator inlet pressure increase breathing loads and cause flow limitation in users of the CRU-103 regulator and MBU-23/P mask?
Does low regulator inlet pressure increase breathing loads and cause flow limitation in users of the CRU-103 regulator and MBU-23/P mask?
Low supply pressures to the CRU-103 regulator cause flow limitations and increased breathing effort, highlighting potential physiological risks during aircraft operations with low engine bleed air pressure.
The life support system in a tactical aircraft provides necessary supplemental oxygen to the aircrew. However, interactions among its various components may generate unexpected breathing loads. We focus here on the interactions between a regulator and breathing mask commonly used together in the U.S. Navy, the CRU-103 regulator and MBU 23/P mask, and some effects of the interactions on the user. The data reported were collected during a larger research effort examining potential physiological and cognitive effects of low regulator inlet pressures. Seventeen participants completed a series of tasks under mild exercise while breathing 40% O2(balance N2) from an MBU-23/P mask supplied by a CRU-103 regulator with supply pressures 10, 6, 4, and 2 psig (CRU-103 specifications are for inlet pressures from 5 to 120 psig). Variables measured included flow to the mask and pressures at the regulator supply, in the hose to the mask, and in the mask. In addition to restricting inspiratory flow, low inlet pressure to the CRU-103 caused a counterintuitive overshoot in gas delivery pressure at end-inspiration, a mean increase of 1.5 cm H2O between the 10- and 2 psig conditions. The added pressure to the exhalation valve increased the expiratory threshold, the pressure to start expiratory flow, by approximately 2 cm H2O, increasing the effort needed to exhale.
No takes yet. Share an insight, caveat, or question.
Robinson et al. (2022) studied Healthy (n=17). Low regulator inlet pressure vs. 10 psig (baseline) was evaluated on Maximum regulator outlet pressure (regulator overshoot) (mean increase of 1.64 cm H2O, p=<0.001).