Key result
A 300 Hz low-pass filter improved the detection of atrial-paced spikes on bedside ECGs compared to the standard 150 Hz filter.
Why the study?
The standard 150 Hz upper-frequency cutoff can cause pacemaker spike loss and misdiagnosis, raising the question of whether alternative low-pass filters improve spike detection on remote and bedside ECGs.
Does a low-pass filter of 300 Hz improve the detection of pacemaker spikes on remote and bedside ECGs compared to the standard 150 Hz cutoff in patients with permanent pacemakers?
Cross-Sectional (n=88)
Does a low-pass filter of 300 Hz improve the detection of pacemaker spikes on remote and bedside ECGs compared to the standard 150 Hz cutoff in patients with permanent pacemakers?
A low-pass filter of 300 Hz significantly improves the detection of atrial and ventricular pacemaker spikes on both bedside and remote ECGs compared to the standard 150 Hz cutoff.
Answers are grounded in this paper and cite their sources.
May improve atrial spike detection on ECGs; hypothesis-generating and requires prospective validation before practice change.
BACKGROUND: The current upper-frequency cutoff of 150 Hz sometimes causes loss of pacemaker spike and misdiagnosis. We hypothesized that low-pass filter (LPF) other than 150 Hz could improve the detection of pacemaker spike. This study aimed to examine the effect of different LPF on pacemaker spike detection in remote and bedside electrocardiogram (ECG). METHODS: Patients with permanent pacemaker implantation were included during routine follow-up. Standard 12-lead ECGs at 6 different upper-frequency cutoff (40, 100, 150, 200, 300, and 400 Hz) were collected. All ECGs were then transmitted to the remote clinic center. Ventricular and atrial pacing were analyzed by 2 independent medical practitioners. RESULTS: A total of 88 patients' ECGs were analyzed (mean age 73.8 ± 10.2 years and 85 with dual-chamber pacemakers). About 75.3% (64/85) of patients were diagnosed as atrial pacing by pacemaker programming. Among 6 different upper-frequency cutoff, the 300 Hz turned out to perform best in detecting atrial-paced spike (area under the curve [AUC] = 0.73, 95% confidence interval [CI]: 0.61-0.84 vs. 0.56, 95% CI: 0.61-0.84 at 150 Hz; P = 0.002) on bedside ECGs. Using programming as the golden standard, the 300 Hz LPF has a sensitivity of 59.4%, specificity of 85.7%, positive predictive value of 92.7% and negative predictive value of 40.9% on bedside ECGs. As for the ventricular pacing, the 300 Hz LPF also had a higher accuracy (AUC = 0.93; 95% CI = 0.84-1.00) than that at 150 Hz (AUC = 0.86; 95% CI: 0.77-0.94; P < 0.001) in detecting ventricular-paced spike on bedside ECGs. The results of remote ECGs were similar with bedside ECGs. CONCLUSIONS: A filter of 300 Hz cutoff may be recommended for ECG spike detection. With the recommended parameter, remote ECG can perform as well as bedside ECG.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2019) conducted a cross-sectional in Permanent pacemaker implantation (n=88). 300 Hz low-pass filter vs. 150 Hz low-pass filter was evaluated on Detection of atrial-paced spike on bedside ECGs (AUC) (95% CI 0.61-0.84, p=0.002). A 300 Hz low-pass filter improved the detection of atrial-paced spikes on bedside ECGs compared to the standard 150 Hz filter.