Answers are grounded in this paper and cite their sources.
Instrumentation study demonstrates a flange-mounted radiofrequency oscillator for mass spectrometers, indicating a compact, low-cost approach for driving nonmass-selective ion guides.
High Resolution Image Download MS PowerPoint Slide In this communication we report the construction of a printed circuit board which mounts directly to the vacuum chamber of a mass spectrometer and produces the RF waveforms needed by many nonmass-selective devices such as ion guides and ion funnels. Our device is designed to replace a standard KF40 flange, can maintain vacuum chamber pressures of less than 10 –6 Torr, and contains the circuitry of the open-source Wisconsin Oscillator RF power supply to generate RF waveforms of 1–4 MHz and up to 200 V p-p . In this iteration of the Wisconsin Oscillator, we also introduce a variable resistor to control the output RF amplitude and show that its ion transmission capabilities are identical to those provided by commercial RF power supplies. With this new implementation we have greatly reduced the space and monetary requirements for driving nonmass-selective ion manipulation devices, which we expect to be advantageous to those developing low-cost and/or portable mass spectrometry systems.
No takes yet. Share an insight, caveat, or question.
Clowers et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: