Key result
The fully automated RNNCNN framework estimated myocardial end-systolic circumferential strain with mean errors of -0.001 to 0.004 compared to manual analysis, processing ~13 slices per second.
Why the study?
The study was designed to demonstrate the feasibility and performance of a fully automated deep learning framework to estimate myocardial strain from short-axis cardiac MRI–tagged images.
Does a fully automated deep learning framework accurately and rapidly estimate myocardial strain from cardiac MRI-tagged images compared to manual analysis?
Cross-Sectional (n=4,508)
Does a fully automated deep learning framework accurately and rapidly estimate myocardial strain from cardiac MRI-tagged images compared to manual analysis?
A fully automated deep learning framework can accurately and rapidly estimate myocardial strain from cardiac MRI-tagged images, enabling high-throughput unbiased evaluation.
May accelerate large-scale tagged CMR strain research; leaves open prospective clinical validation before routine use.
Purpose To demonstrate the feasibility and performance of a fully automated deep learning framework to estimate myocardial strain from short-axis cardiac MRI–tagged images. Materials and Methods In this retrospective cross-sectional study, 4508 cases from the U.K. Biobank were split randomly into 3244 training cases, 812 validation cases, and 452 test cases. Ground truth myocardial landmarks were defined and tracked by manual initialization and correction of deformable image registration using previously validated software with five readers. The fully automatic framework consisted of (a) a convolutional neural network (CNN) for localization and (b) a combination of a recurrent neural network (RNN) and a CNN to detect and track the myocardial landmarks through the image sequence for each slice. Radial and circumferential strain were then calculated from the motion of the landmarks and averaged on a slice basis. Results Within the test set, myocardial end-systolic circumferential Green strain errors were −0.001 ± 0.025, −0.001 ± 0.021, and 0.004 ± 0.035 in the basal, mid-, and apical slices, respectively (mean ± standard deviation of differences between predicted and manual strain). The framework reproduced significant reductions in circumferential strain in participants with diabetes, hypertensive participants, and participants with a previous heart attack. Typical processing time was approximately 260 frames (approximately 13 slices) per second on a GPU with 12 GB RAM compared with 6–8 minutes per slice for the manual analysis. Conclusion The fully automated combined RNN and CNN framework for analysis of myocardial strain enabled unbiased strain evaluation in a high-throughput workflow, with similar ability to distinguish impairment due to diabetes, hypertension, and previous heart attack. Keywords: Adults, Cardiac, MR-Imaging, Neural Networks Published under a CC BY 4.0 license. Supplemental material is available for this article.
No takes yet. Share an insight, caveat, or question.
A 2020 study conducted a cross-sectional in Cardiovascular risk factors (UK Biobank cohort) (n=4,508). Fully automated deep learning framework (RNNCNN) vs. Manual analysis was evaluated on Myocardial end-systolic circumferential Green strain error. The fully automated RNNCNN framework estimated myocardial end-systolic circumferential strain with mean errors of -0.001 to 0.004 compared to manual analysis, processing ~13 slices per second.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: