Metallurgical and histological examinations of implants and adjacent tissue removed from orthopedic patients have been performed in a series of 190 cases. The results have been correlated with clinical findings where possible. The two major factors to emerge from this study were that many stainless steel implants corrode in the body and that titanium implants release titanium into the local tissue, which may become discolored. Severe cases of corrosion in stainless steel implants were attributed to faulty manufacturing technique and poor implant usage. This corrosion was frequently associated with crystal‐like “microplates” in the tissue and lead to clinically significant effects in some cases. Less severe cases of corrosion were almost invariably associated with interfaces between components. Susceptibility to this corrosion may be minimised by improvements in design and materials specification. Titanium levels in excess of 2,000 ppm were found in the tissue in 3 cases out of 19 analysed, no specimen of a control series showing greater than 100 ppm. Histological sections showed accumulations of intracellular particles of exogenous origin. The evidence that such titanium release and particle accumulation causes clinically significant effects is only minimal but it is possible that systemic distribution of titanium and remote hypersensitivity effects could arise.
No takes yet. Share an insight, caveat, or question.
Williams et al. (1974) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: