Answers are grounded in this paper and cite their sources.
Orthopedic appliances of type 316 (SMO) stainless steel were examined in detail after being removed from human subjects. Techniques included conventional microscopy with various lighting methods, together with electron microscope fractography. Corrosion, fatigue failure, and manufacturing defects were identified and statistically tabulated. Corrosion damage occurred on 91% of all multicomponent devices and the overall incidence of fracture failure was 9%. These results indicate an urgent ueed for stronger, more corrosion‐resistant implant alloys.
No takes yet. Share an insight, caveat, or question.
Colangelo et al. (1969) studied this question.